RA3100 Remote Control Module

TRIG IN/OUT.

For use with a remote control module cable to connect with other devices (available separately).

Remote Control Module RA30-112 Specifications
Input Connector Half-pitch 20-pin connector
Output Connector Half-pitch 14-pin connector
External Input
Function: Control by external signal.
Control Signal START/STOP, MARK, FEED, PRINT, TRIG
Input Level High level: 2.1V to 5.0V
Low level: 0V to 0.5V (active low)
Response Speed Select from High-speed/Normal/Low-speed
Effective Pulse Width High-speed response: 1μs or higher during high interval, 1μs or higher during low interval
Normal response: High interval 1ms or higher, Low interval 1ms or higher
Low-speed response: 10ms or higher during high, 10ms or higher during low-speed response
Maximum Allowable Input Voltage 30V
External Output
Function: Control signals can be externally output
Control Signal START/STOP, MARK, FEED, PRINT, TRIG
Output Level High level: 3.8V to 5.0V
Low level: 0V to 0.5V (active low)
Output Pulse Width START/STOP, FEED, PRINT:
Active output during operation
TRIG, MARK:
High-speed response: 1 μs
Normal response: 1 ms
Low-speed response: 10 ms
External Sampling Input (EXT.SMPL IN)
Synchronization via external clock signal is possible (simultaneous SSD recording and printer recording are not possible.)
Input Level High level: 2.1V to 5.0V
Low level: 0V to 0.5V
Effective Pulse Width High-speed (SSD Recording): 2μs or higher
Low-speed (Printer Recording):1ms or higher
Maximum Input Frequency High-speed (SSD Recording): 250 kHz
Low-speed (printer recording): 500Hz
External Sampling Output (EXT.SMPL OUT)
Synchronization clock signal can be output externally
Output Level High level: 3.8V to 5.0V
Low level: 0V to 0.5V (active low)
Reference Clock for Calibration
Function: Clock output for operation check
Output Level 0V to 5V (±1%)
Output Frequency 1kHz (±1%)
Duty Ratio 50% (±5%)
General
Withstand Voltage AC300V, 1 minute (between input/output and main chassis)
Maximum Rated Voltage to Ground AC, DC 42V
Dimensions Approx. 140 (input-side W) x 223(D) x 20(H) mm
Weight Approx. 250g
Compliance Standards Safety: IEC61010-1
EMC: IEC61326-1, class A