UK Supplier of Data Acquisition & Thermal Imaging systems, software and services

Search

RA3100 Multi-channel, Portable Thermal Chart Recorder

Multi-channel input – Max 36ch (analogue input), Max 144 points (digital input).

High-speed sampling (Max 20MS / s).

A&D 3100 Chart Recorder Trade In Discount Promotion

Manufacturer:

£5,655.00 ex VAT

Available on back-order

Description

RA3100 Thermal Chart Recorder / Portable DAQ System from A&D.

The RA3100 accurately produces high-speed switching waveforms even in harsh noise environments at research and development sites and maintenance sites for electric vehicles, home appliances, railroad vehicles, solar power generation systems, etc. that use inverter control technology. 

Product Features

  • Multi-channel input
    Max 36ch (analogue input), Max 144 points (digital input)
  • High-speed sampling
    Max 20MS / s
  • Long-term recording
    Memory capacity 4GB (18ch / 20ms / s, 5.6 seconds), SSD capacity 256GB (36MS, 1ms / s, approx. 50 minutes)
  • Input unit
    Voltage, temperature, logic
  • High-speed and High-definition printing
    Maximum speed 100 mm / s
    Backup to SSD even if there is no chart paper
  • Excellent visibility and operability
    Better visibility and operability with the adoption of a 12.1-inch LCD touch panel.
  • Back scroll
    Data being measured can be played back without ending the measurement
  • Various monitors
    YT waveform display, XY display, FFT analysis can be displayed on the LCD monitor during measurement

 

Module Options for RA3100 Thermal Chart Recorder / Portable DAQ System.

RA3100 Modules Range

Module
Sampling Rate
Input
Specifications
2ch Voltage 1MS/s Voltage ±500V Measure high-speed voltage measurement with anti-aliasing filters
4ch Voltage 1MS/s Voltage ±200V Multi-channel voltage measurement
2ch High-speed Voltage 20MS/s Voltage ±500V High-speed voltage measurement
2ch High Voltage 1MS/s Voltage ±1,000V Module to measure high voltages of ±1,000V
16ch Logic 1MS/s Contact, Voltage Contact, Voltage signal measurement
2ch Temperature 1.5ms Thermocouple : K, E, J, T, N, R, S, B, C
RTD: Pt100, Pt1000
Measurement of temperature with a thermocouple and RTD
2ch AC Strain 100kS/s Strain gauge, Strain gauge transducer Measures stress, load, displacement, pressure, torque, and acceleration
2ch Acceleration 1MS/s Piezoelectric acceleration transducer
(charge output, built-in amplifier)
Measures acceleration, speed and displacement
2ch Frequency 1MS/s Voltage ±500V Input module capable of measuring period, rotation speed, number of pulses, etc., of input signals

Available Separately. For more information visit the product pages via the links below:

2ch Voltage Module
4ch Voltage Module 
2ch High-speed Voltage Module 
2ch High Voltage Module
16-channel Logic Module 

2ch Temperature Module 
2ch Frequency Module
2ch AC Strain Module
2ch Acceleration Module
Remote Control Module 

 

RA3100 Specifications

Main Specifications
Display 12.1-inch TFT colour liquid crystal display (1024 x 768), with capacitive touch panel
Channel Number of module slots: 9
Analogue measurement: Up to 36 channels (when using 4ch voltage module)
Logic measurement: Up to 144 channels (when using 16ch logic module)
Memory Capacity 4GB (2G points/ch)
Memory Division 1, 2, 4, 8, 16, 32, 64, 128 split
Sampling Speed Memory recording: 20 MS / s (50 ns) max. to 10S/min
SSD recording: 1 MS / s (1 μs) max. to 10S/min
Storage Device SSD (Solid State Drive) 256GB
SD card slot, SD / SDHC / SDXC compatible, for storing data after recording
USB memory using USB port, for storing data after recording
Printer Thermal recording with thermal head
Paper width: 219.5mm
Chart Speed: 100mm / s to 1mm / min
Effective recording width: Maximum 215mm
Printer Functions

Memory Recording: Record high-speed event recording to memory * 
SSD Recording: Recording of the input signal to the internal SSD
Printer Recording: Recording of the input signal directly to a thermal printer

*Any combination of memory recording, SSD recording, and printer recording is possible.

 Trigger Uses: Trigger for start recording operations, trigger for memory recording
Source Channels: Analog/logic signals, manual triggers and external signal triggers of each module
Types: Level triggered, window triggered (memory recording triggered ), bit pattern triggered
Trigger Filters: Level hold, count
Trigger Output: Output signal when trigger conditions are met
Monitor Y-T Waveform Monitor: Display time changes for a measurement signal in a pulse amplitude waveform
X-Y Waveform Monitor: Data is plotted with common points, with Signal 1 as the X axis and Signal 2 as the Y axis
FFT Analysis Monitor - FFT analysis of the measurement signal of any two channels is performed, and the analysis results are displayed as the frequency axis data.
Recording and Display
Y-T Waveform Displayed Graphs: 1 graph
Sheets: Measurement channels are displayed in three sheets, and 48 channels are registered per sheet.
Grid: 20 div vertically and 20 div horizontally
Time axis Data Count: 100 data/div
Chart Paper Resolution: Pulse amplitude axis 8dot/mm, time axis 80dot/mm (at 25mm/s) 
X-Y Waveform Data Selection: Select from input signal or recorded data
Displayed Graphs: 1 graph, 4 graphs
Grid: 20 div vertically and 20 div horizontally
Sampling Speed: Max 1KS/s
Locas: ON/OFF of locas enabled (pen up& down)
Printing: Print the plotted X-Y waveform with the printer
FFT Analysis Data Selection: Select from input signal or recorded data
Sampling Points: 1,000, 2,000, 5,000, or 10,000 points
Max Analysis Frequency: 1/2.5 times the sampling frequency
Displayed Graphs: 1 graph, 2 graphs. The Y-T waveform can also be displayed
Graph Types: Waveform in time axis , linear spectrum, RMS spectrum, power spectrum, 1/1 octave, 1/3 octave, Transfer function, cross power spectrum, and coherence function
Window Function: Hanning Window
Average Processing: Time axis simple addition average, frequency axis simple addition average, frequency axis exponentially weighted average, frequency axis peak hold
Interface
LAN 1000BASE-T (1Gbps): For control with communication command
COM RS-232C: For control by communication command or for UPS shutdown
USB Ver. 3.0 2 port: For storage devices (USB memory)
Video Output DVI-D
Operating Environment Temperature 5 to 40°C
Humidity 35 to 80% RH (without condensation)
Power Supply 100 to 240VAC, Frequency 50 to 60Hz
Power consumption 300VA or less (under the maximum load conditions)
Dimensions 394 (W) × 336(H) × 200(D) mm *Excluding projections
Weight 9.5kg or less (main body only)
11 kg or less (when main body and four 2-channel voltage modules are installed)

 

Module Specifications

Available separately.

2ch Voltage Module

No of channels 2
Input Coupling AC, DC, and GND coupling
Maximum Sampling Speed 1MS/s
Frequency Response (-3dB) DC to 100kHz
A/D Resolution 16bit
Range Accuracy ±0.3% of RANGE *After offset cancellation
Allowable Input Voltage ±500V (DC or AC peak values)
Measurement Range 100mV to 500V (1-2-5 step)
Input Terminals and Models Insulated BNC connector, insulated unbalanced input, isolation between each channel
Input Impedance 1MΩ
Low-Pass Filter 3, 30, 300, 3k, OFF
Anti-Aliasing Filter 10 to 40kHz
Maximum Rated Voltage To Earth AC, DC300V

4ch Voltage Module 

No of Channels 4
Input Coupling DC and GND coupling
Maximum Sampling Speed 1MS/s
Frequency Response (-3dB) DC to 100kHz
A/D Resolution 16bit
Range Accuracy ±0.2% of RANGE *After offset cancellation
Allowable Input Voltage ±200V (DC or AC peak values)
Measurement Range 1V to 200V (1-2-5 step)
Input Terminals and Models Insulated BNC connector, insulated unbalanced input, isolation between each channel
Input Impedance 1MΩ
Low-Pass Filter 3, 30, 300, 3k, OFF
Maximum Rated Voltage To Earth AC, DC300V

2ch High-speed Voltage Module 

No of Channels  2
Input Coupling AC, DC, and GND coupling
Maximum Sampling Speed 20MS/s
Frequency Response (-3dB) DC to 5MHz
A/D Resolution 14bit
Range Accuracy ±0.5% of RANGE *After offset cancellation
Allowable Input Voltage ±500V (DC or AC peak values)
Measurement Range 100mV to 500V (1-2-5 step)
Input Terminals and Models Insulated BNC connector, insulated unbalanced input, isolation between each channel
Input Impedance 1MΩ
Low-Pass Filter 5, 50k, 500k, OFF
Maximum Rated Voltage To Earth AC, DC300V

2ch Temperature Module 

No of Channels 2
Maximum Sampling Speed 1kS/s
A/D Resolution 16bit
Minimum Resolution 0.1°C for the thermocouple and 0.01°C for RTD
Thermocouple Types K, E, J, T, N, R, S, B, C
RTD Types Pt100, Pt1000
Range Accuracy ± (0.1% of RANGE + 1°C) *When 0°C or higher and using K, E, J, T, or N
Cold Connection Point Compensation Accuracy ±1℃
Input Terminals and Models Spring type terminal block, insulated unbalanced input, isolation between each channel
Maximum Rated Voltage To Earth AC, DC300V

16-channel Logic Module 

Bit 16bit
Maximum Sampling Speed 1MS/s
Input Type Voltage or contact, isolated input, no insulation between channels
Input Voltage Range 0~24V
Input Response Time 1μs
Input Terminal 8ch square connector x 2 ports
Notes Input by an optional IC clip or electrical clip code

Remote Control Module 

Features Start /Stop by external signal, mark print, paper-feed, external sample
Input of external trigger signal, trigger signal output
Synchronous digital signal for AC distortion module
RA3100 and Accessories Specifications RA3100 and Accessories Specifications